FEATURE
• Maximum Output 32kW
■ The system can be used at Patient room, Emergency
room, ICU,etc.
■ Touch screen user interface which includes a virtual
keypad to memorize the anatomical techniques and,
in case of D.A.P., entering the patient data; the dose
is displayed directly on the touch screen.
SPECIFICATION
Standard Output Power: 32kW
Frequency: 100 kHz
Type of Inverter: HF1100
The Max Electric Current: 425mA
Monoblock max Voltage : 125Kv
mAs:0.2- 220mAs
Ripple :<3%
Focus Range: 0.6 / 1.3mm
Monoblock Termic capacity: 375 KJ
Monoblock Continuous Thermal Dissipation: 40W
Variation KV: Steps 1KV
Operating Methods : two-points radiological tech¬
nique with selection of kV/mAs, or three points
radiological technique with selection of kV/mA/mS
[optional]. The values and the exposure time are
always selected and displayed, automatically by
the microprocessor, on the touch screen display
single-phase 230Vac. Maximum consumption: 12A.
Any power socket can be used. In fact, the system
operates with a capacitor bank to store energy.
By touch screen you can select the two charging
levels of the capacitor bank. The first level allows
most radiographic to be carried out with a reduced
charge of the capacitors. The second level is only
selected